A&D UT-302PBT Thermometer User Manual


 
20
EMC Tables
Recommended separation distances between portable and mobile RF communications
equipment and the Thermometer
The monitor is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
as recommended below, according to the maximum output power of
the communications equipment.
Thermometer
Thermometer
Thermometer
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHZ
d=1.2P
80 MHz to 800 MHZ
d=1.2P
800 MHz to 2.5 GHz
d=1.2P
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where Pis the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
These guidelines may not apply inall situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
NOTE 1.
NOTE 2.
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
UT-302PBT
UT-302PBT
UT-302PBT
UT-302PBT