A&D UT-302PBT Thermometer User Manual


 
18
EMC Tables
Guidance and manufacturer's declaration-electromagnetic emission
The intended for use in the electromagnetic environment specified
below.The customer or the user of the should assure that it is used
insuch an environment.
Thermometer
Thermometer
Emissions test Compliance Electromagnetic environment guidance
RF emissions CISPR 11 Group 1The uses RF energy only
for its internal function.Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
Thermometer
RF emissions
CISPR 11
Class B
The is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Thermometer
Harmonic
emissions
IEC 61000-3-2
Not
applicable
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
Guidance and manufacturer's declaration-electromagnetic immunity
The is intended for use in the electromagnetic environment specified
below.The customer or the user of the should assure that it is used
insuch an environment.
Thermometer
Thermometer
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
3A/m 3A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical.
UT-302PBT is
UT-302PBT
UT-302PBT
UT-302PBT
UT-302PBT
UT-302PBT