A&D UT-302PBT Thermometer User Manual


 
19
EMC Tables
Guidance and manufacturer's declaration-electromagnetic immunity
The is intended for use in the electromagnetic environment
specifiedbelow. The customer or the user of the should assure that
it is
used in such an environment.
Thermometer
Thermometer
Immunity
test
IEC 60601 test
level
Compliance
l
evel
Electromagnetic environment-guidance
Radiated
RF IEC
61000-4-3
3 V/m
80 MHz to 2,5
GHz
3 V/mPortable and mobile RF communications equipment
should be used no closer to any part of the
, including cables, than the
recommended separationdistance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d=1.2P
d=1.2P80 MHz to 800 MHZ
d=2.3P800 MHz to 2,5 Ghz
where Pis the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey-a,
should be less than the compliance level in each
frequency range-b.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Thermometer
NOTE 1.
NOTE 2.
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection
from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast andTV
broadcast cannot be predicted theoretically with accuracy.To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the
is used exceeds the applicable RF compliance level above, the
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.