A&D UA-767 Blood Pressure Monitor User Manual


 
ENGLISH 6
Measurements
1. Place the cuff on the arm (preferably the left arm).
2. Set the PRESSURE switch to a value that will be 30 to 40 mmHg above the
expected systolic value and press the START button.
When the START button is pressed, all of the display symbols will appear
for about one second.
When the " 0 " is flashing, the device is calibrating its zero point. If there is
still some air in the arm cuff, it will be exhausted at this time.
The micro pump will then pressurize the cuff.
Note
The cuff will become very tight. This is normal.
If the START button is pressed again during pressurization or during the
measurement, the pump will stop and the air will be released from the cuff.
3. When pressurization is complete, the automatic exhaust mechanism will
gradually reduce the pressure in the cuff and the Measurement in Progress
mark will appear along with the current pressure reading. This mark will
flash in unison with the pulse of the person being monitored.