A&D HV-15KV-WP Scale User Manual


 
Page 50 HV-WP/HW-WP Series
16. RS-232C Serial Interface
16.2.
16.2.16.2.
16.2. Stream Mode
Stream ModeStream Mode
Stream Mode
The scale outputs the current display data at the time when refreshing the display.
The scale does not output data while in the setting mode.
Averaging of function table Refresh rate
A
f13 0 Normal weighing
Approximately 7 times/s while the display is unstable,
Approximately 4 times/s when the display is stable
A
f13 1 A person’s weight
A
f13 1 Animal weighing
Approximately 4 times/s
A
16.2.1.
16.2.1.16.2.1.
16.2.1. Preparation and Performance (Examples)
Preparation and Performance (Examples)Preparation and Performance (Examples)
Preparation and Performance (Examples)
Step 1 Turn off the display.
Press the ON/OFF switch while the ZERO switch is
pressed and held. The function table is displayed.
Step 2 Select a parameter for the "baud rate ( f4 )" using
the ENTER and switches.
Step 3 Select "stream mode ( f5 0 )" of "output" using the
ENTER and switches.
Step 4 Select "format 1 ( f15 0
)" of "format" using the
ENTER and switches.
Step 5 Press the ENTER switch to store the new
parameters. The scale returns the normal weighing
and outputs the data coutinuouslly.