Remote Commands
STATus Commands
Figure 8-2 The Status Registers
:STATus:OPERation:CONDition?
Syntax :STATus:OPERation:CONDition?
Description This query reads the contents of the
OPERation:CONDition register. Only three
bits of the condition register are used:
Bit 1, which is 1 when the motor that
positions the attenuator filter is settling.
Bit 3, which is 1 while the instrument is
performing an attenuation sweep.