Varec 4532 Thermometer User Manual


 
Operation and Description of Instrument Function Average Temperature/Water Bottom Sensor and Converter
44 Installation and Operations Manual
VH92 Error Dis. Sel.
Item type: select
Default: 0_OFF
Selection: 0_OFF, 1_ON
Type of VH88 "Short Error Value" and VH89 "Open Error Value" display selection.
0_OFF: These 2 error messages will not be transmitted to the connected host gauge.
This function automatically excludes defect element in average temperature calculation.
1_ON: Error message will be transmitted to the host gauge. As a result, VH88 and 89"s
numeric error code will be displayed on host gauge default screen and may transmit to
upper receiver as well.
VH94 Polling Address
Item type: read and write
Default: 2
Range: 1 ~ 15
Polling address for HART communication
VH95 Manufacture ID
Item type: read only
Default: 17
The manufacturer ID for Endress+Hauser instrumentation.
VH96 Software Version:
Item type: read only
A display of installed software version.
VH97 Hardware Version
Item type: read only
A display of installed hardware version.
VH98 Below Bottom
Item type: select
Default: 0_OFF
Selection: 0_OFF, 1_ON
A type of error display when liquid level drops below #1 element (Bottom Point). Error
code "29" is displayed on VH80 and VH91 when 0_ON is selected.
VH99 Device Type Code
Item type: read only
Device type will be displayed.
190: Temperature measurement function only.