Varec 4532 Thermometer User Manual


 
Varec, Inc. 43
4532 Operation and Description of Instrument Function
VH87 Element Interval
Item type: read and write
Default: 1000mm
Range: 0mm ~ 99999mm
Applies only when Even Interval spacing is selected.
Caution! Changing element interval, setting element position, these are only applied to recon-
figure switching points for average temperature calculation. The physical element
position will never be changed.
VH88 Short Error
Item type: read and write
Default: -49.5
Range: -49.5 ~ 359.5
Error message displayed when any element has a short circuit. Method of display can be
configured at VH92 "Error Display Select."
VH89 Open Error
Item type: read write
Default: 359.9
Range: -49.5 ~ 359.5
Error element displayed when any element has an open circuit. Method of display can be
configured at VH92 "Error Display Select."
8.1.9 Device setting 2: VH90~VH99
VH90 Device ID Number
Item type: read and write
Default: 0
Range: 0 ~ 16777214
In order to distinguish own device ID when 4532 ATC is connected in multi drop HART
loop.
Caution! Changing device ID may lead to communication error because of mismatched pre-
registered device ID and HART address.
VH91 Previous Error
Item type: read only
Display of error history. Coded error message will be the same contents as VH80.