Varec 4532 Thermometer User Manual


 
Varec, Inc. 41
4532 Operation and Description of Instrument Function
VH80: Present Error
Item type: read only
Display of current error message. One of the following codes will be indicated:
Error code
0: No error presence
1: Common line open
2: undetermined
3: #1 element open
4: #1 element short
5: #2 element open
6: #2 element short
7: #3 element open
8: #3 element short
9: #4 element open
10: #4 element short
11: #5 element open
12: #5 element short
13: #6 element open
14: #6 element short
23: #0 element over range
24: Memory defect (ROM)
29: Element exposed (liquid level below #1 element position)
30: undetermined
31: undetermined
41: Memory defect (RAM)
42: Memory defect (EEROM)
VH81 Temperature Unit
Item type: select
Default: °C
Selection: C, F, K
Selection of temperature display unit. Based on universal HART setting, °C(HART code:
32), °F(HART code: 33) and °K(HART code: 35) are available.
Note! Selection of temperature display unit only applies to reply data from 4532 ATC. Data
transmission from host gauge (4590 TSM or 6000 Series Servo Tank Gauge) to
4532 ATC must be performed by °C unit only (terminology of HART command 133)