4. OPERATION
1. Place the crosslinker on a level work surface. Be sure there is enough room in front for
the door to open easily.
2. Plug the female end of the power cord into the crosslinker.
3. Plug the male end of the power cord into a properly grounded electrical outlet. The
proper operating voltage of the crosslinker is found on the product information label.
Note: For 230V models, or those requiring special power cord connectors, ensure that the
proper configuration of the male connector or the plug has been properly connected to the
power cord.
4. Turn the On/Off switch to the ON position. Note: WHEN TURNED ON THE
CROSSLINKER DEFAULTS TO THE LAST USED UV EXPOSURE SETTING.
5. The last UV exposure setting will now be displayed on the LED and the last function
setting will be noted by glowing red spot(s) on the display panel.
6. Place your sample requiring exposure into the chamber.
7. The UV Crosslinker can be operated on the following settings:
A. Preset ultraviolet energy exposure setting
Push the PRESET and then ENERGY on the tactile touchpad. The red light at each
position should now be lit and the preprogrammed UV exposure setting of 120,000
microjoules per cm² displayed in the LED as 1200. Note: THE LED DISPLAYS
1200. THIS MUST BE MULTIPLIED BY 100 TO OBTAIN THE EXPOSURE.
Push START. After a slight delay to energize the UV tubes, the LED will begin to
countdown. The unit automatically stops at the end of the exposure cycle and will
beep FIVE times. Exposure is now complete.
NOTE: Though the PRESET UV EXPOSURE SETTING is factory set at
120,000 microjoules per cm², it is possible to change this if necessary, as follows:
Push and hold the PRESET on the tactile touchpad until you hear an audible signal
then push ENERGY also on the touchpad.
Set your new Preset UV exposure by pushing the numbers on the touchpad.
- 3 -
Remember: THE LED DISPLAY MUST BE MULTIPLIED BY 100 TO OBTAIN
THE PROPER EXPOSURE SETTING.