A.4 Reference Drawing
D-77712 Outline Diagram
D77713 Control Unit Final Assembly
C77925 Interconnection Diagram
A.5 Probe or Sensor EMF Tables
ZIRCONIA OXYGEN SENSOR OUTPUT (mV)
PROBE TYPE 9060L, SENSOR TYPE 9060LEX
% OXYGEN mV at 720°C (1320°F)
21.0 0.00
20.5 0.46
20.0 0.99
19.5 1.53
19.0 2.09
18.5 2.66
18.0 3.25
17.5 3.85
17.0 4.47
16.5 5.11
16.0 5.77
15.5 6.45
15.0 7.15
14.5 7.87
14.0 8.62
13.5 9.40
13.0 10.21
12.5 11.05
12.0 11.92
11.5 12.83