Percent Oxygen Analyzer Maintenance
Teledyne Analytical Instruments 41
Maintenance
Aside from normal cleaning and checking for leaks at the probe
connection, the Model 9060L should not require any maintenance
beyond replacement of a blown fuse. Routine maintenance includes
occasional recalibration, as described in Chapter 4,
Operation.
5.1 Replacing the Fuse
CAUTION: REMOVE POWER TO UNIT BEFORE REPLACING
THE FUSE.
When a fuse blows, check first to determine the cause, then replace
the fuse using the following procedure:
1.
Disconnect the AC power and place the power switch located on
the rear panel in the
O position. Remove the power cord from the
receptacle.
2.
The fuse receptacle is located in the power cord receptacle
assembly in the upper left-hand corner of the rear panel. See
Figure 5-1.
Figure 5-1: AC Fuse Replacement