Teledyne 3000TA-XL-EU Oxygen Equipment User Manual


 
Trace Oxygen Analyzer Installation 3
3-1
Teledyne Analytical Instruments
Installation
Installation of the Model 3000TA-XL Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Installing the Micro-Fuel Cell
6. Testing the system.
3.1 Unpacking the Analyzer
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the analyzer and inspect
it for damage. Immediately report any damage to the shipping agent.
3.2 Mounting the Analyzer
The Model 3000TA-XL is for indoor use in a general purpose area. It
is NOT for hazardous environments of any type.
The standard model is designed for flush panel mounting. Figure 3-1 is
an illustration of the 3000TA-XL standard front panel and mounting bezel.
There are four mounting holes—one in each corner of the rigid frame.
Drawing number D-62928, at the back of this manual, contains a panel
cutout diagram.
On special order, a 19" rack-mounting panel can be provided. For rack
mounting, one or two 3000XL series analyzers can be flush-panel mounted
on the rack panel. See Appendix for dimensions of the mounting panel.