Disassembly/Reassembly, and Adjustment (cont)
Disassembly
Access to Motor Brushes
a. For quick access to Motor Brushes, perform steps 6b of
this section.
b. Use a flat blade screwdriver to remove the
motor brush cap and then remove the brushes for inspec-
tion.
(figure A4.21).
c. Inspect Motor brushes for excessive wear, uneven wear,
and/or severe discoloration.
d. If any of these conditions exist, replace/install both
brushes for the motor in question.
Hint: Make sure that the brushes are seated properly
before re-installing the brush caps.
Reassembly
a. Reassemble your S11, by performing the Reverse of
steps 6b.
Step 8 - Check / Replace Motor Brushes