Sunrise Medical S-11 Mobility Aid User Manual


 
QUICKIE S-11 SERVICE MANUAL
PAGE 16
2005SUNRISE MEDICAL
Disassembly
Note: For ease of disassembly, set the base frame on a
block where all four wheels are at least 1 inch above the
ground.
a. Unwrap the wire looms from the motor and pull the
motor connector out (figure A4.17).
b. Depress the motor connector locking tab and unplug it.
c. Straighten the lock washer tab of the drive wheel.
d. Use a 13mm socket wrench to remove the drive
wheel retention bolt, then pull the drive wheel out from the
motor shaft (figure A4.18).
e. Use a 5mm Allen wrench to remove the 4 mounting
screws
(figure A4.19).
f. Pull the motor-wheel assembly out .
Note: If the drive wheel is difficult to remove, then remove
the wheel plate screws (5mm Allen head)
Reassembly
a. Perform the reverse of instructions above.
Note: Torque specifications
* Motor mounting screws 15-20 ft-lbs
* Wheel retention bolt 15-20 ft-lbs
* Wheel plate screw 15-20 ft-lbs
Adjustment
Note: There is a suspension bumper on the frame that lim-
its
the top frame’s movement (figure A4.20).
a. Use a 13mm open wrench to loosen or tighten the jam
nut and make the proper adjustment (operator preference).
Step 7 - Motor/Wheel
Disassembly/Reassembly, and Adjustment (cont)
A4.17
A4.19
A4.20
A4.18