ResMed S7 Respiratory Product User Manual


 
28
Recommended separation distances between portable and mobile RF communications
equipment and the ResMed S7 Lightweight
The ResMed S7 Lightweight is intended for use in an environment in which radiated RF disturbances are
controlled. The customer or the user of the ResMed S7 Lightweight can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the ResMed S7 Lightweight as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output
power of transmitter
W
150kHz to 80MHz
d = 1.17 P
80 MHz to 800 MHz
d = 0.35 P
800MHz to 2.5 GHz
d = 0.35 P
0.01 0.17 0.04 0.04
0.1 0.37 0.11 0.11
1 1.17 0.35 0.35
10 3.69 1.11 1.11
100 11.70 3.50 3.50
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.