SYSTEM SPECIFICATIONS
GUIDANCE AND MANUFACTURERS DECLARATION - ELECTROMAGNETIC EMISSIONS
AND IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic emissions
The ResMed S7 Lightweight is intended for use in the electromagnetic environment specified below. The
customer or the user of the ResMed S7 Lightweight should assure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment -
guidance
RF emissions CISPR11 Group 1 The ResMed S7 Lightweight uses RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11 Class B The ResMed S7 Lightweight is suitable
for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage network that supplies buildings
used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker
Emissions IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to EMC information provided in this document.
Warnings: The ResMed S7 Lightweight should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the ResMed S7 Lightweight should be observed to verify normal
operation in the configuration in which it will be used.
The use of accessories (eg Humidifiers) other than those specified in this manual is not recommended.
They may result in increased emissions or decreased immunity of the ResMed S7 Lightweight.