ResMed Hospital Full Face Mask Respiratory Product User Manual


 
ENGLISH
The exhaust vent flow is lower at lower mask pressures (see Pressure/Flow Curve). If the
mask pressure during expiration is too low, there may not be enough exhaust vent flow
to completely flush the air delivery hose of exhaled air before the start of the next
inspiration.
Discard the mask if heavily soiled (eg, gastric contents).
If oxygen is used with the CPAP or bilevel device, the oxygen flow must be turned off
when the flow generator is not operating.
Explanation of the Warning: When the CPAP or bilevel device is not in operation, and
the oxygen flow is left on, oxygen delivered into the ventilator tubing may accumulate
within the flow generator enclosure. Oxygen accumulated in the CPAP or bilevel device
enclosure will create a risk of fire. This warning applies to most types of CPAP or bilevel
flow generators.
Please refrain from smoking while oxygen is in use.
Note: At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration will
vary, depending on the pressure settings, patient breathing pattern, mask selection, and the leak
rate.
FITTING THE MASK
The RESMED HOSPITAL FULL FACE MASK is supplied fully assembled.
Note: Inspect the mask and replace it if any parts are damaged.
1 Connect one end of the main air tubing to the flow generator and the other end to the
swivel. Turn the flow generator on. See “Technical Specifications” for information on the
correct flow generator setting.
2 Place the mask cushion over the patient’s nose and mouth and pull the headgear over
their head.
3 Gently tighten the headgear straps:
a first, pull the side straps back along the side of the patient’s face.
b then, adjust the top strap.
headgear
Back view Top view
60867r3.book Page 2 Thursday, December 14, 2006 4:56 PM