Princeton 4411-0019 Home Dialysis Equipment User Manual


 
7
Chapter 2
Warnings
Introduction
This chapter contains cautions and notes concerning initial setup and operation of your
system. Of course, reading this section should not replace reading the entire manual. The
rest of the manual contains complete procedures for installing and operating your
system, as well as a more detailed explanation of the following warnings
Verify that the line voltage and fuse ratings are compatible before turning
on the system.
Voltage applied outside the range may damage the apparatus! The line
voltage is not selectable. The unit must be modified at the factory for the desired
operating voltage. The Model number on the rear panel indicates the line voltage with
which this unit is designed to operate. The Model CC100-115V may be used for line
voltages between 105 and 125 V AC, 50 or 60 Hz. The Model CC100-230 V may be
used for line voltages between 210 and 250 V AC, 50 or 60 Hz. The Model CC100-100V
may be used for line voltages between 90 and 100 V AC, 50 to 60 Hz. The fuse ratings
are labeled on the rear panel, except Model CC100-100V. For the CC100-100V the fuse
rating is 2 A at 120V AC.
Protective grounding must be maintained.
Protective grounding could become
ineffective in apparatus possibly damaged during shipping. Until safety has been verified
by Princeton Instruments, Inc, such damaged apparatus should not be used. Tag the unit
appropriately to warn any potential user of such a danger.
A socket outlet provided with the necessary earth ground contact must be
used.
Any extension cord or adapters used must also comply with these safety
requirements. If the supplied cord plug is not compatible with the power socket, a
substitute plug of a compatible design should replace it. Replacement power cords or
power plugs must have the same polarity as those of the original equipment to avoid
hazard due to electrical shock. Disconnecting the protective earth terminal inside or
outside the apparatus is prohibited.
Never run the CC-100 without coolant in the reservoir.
Although the magnetic
pump can withstand short intervals of operation without coolant, doing so will shorten
the life of the pump.
WARNING
WARNING
WARNING
CAUTION