6 CC-100 Coolant Circulator Manual Version 1.B
Cooling Capacity
Temperature rises approximately 0.03°C per watt (3.4 btu/hr). Based on a typical
flow rate of between 0.5 (0.47 liter) to 1 quart (0.95 liter) per minute and a thermal
load of 100 watts (341 btu/hr).
When using a CC-100 to cool a PI detector, the coldest attainable temperature should
be reduced by no more than two degrees compared with using tap water at ambient
temperature.
Operating Environment
Temperature
Minimum:
0°F (-18°C). Avoid freezing coolant.
Maximum:
120°F (49°C).
Humidity:
Avoid condensation
Clearance
Front:
6 inches (150 mm) minimum
Rear:
12 inches (300 mm) minimum
Side:
1 inch (20 mm) minimum
Power Requirements:
50/60 Hz
2 A @ 105V to 125V AC (standard)
2 A @ 90V to 110 V AC (factory installed option)
1 A @ 210 V to 250 V AC (factory installed option)
Mechanical
Dimensions
D, W, H: 12½ 318mm) × 15¾ (400mm) × 8¼ (210 mm)
Weight (without coolant)
16 lb (7.3 kg) without transformer, standard
19 lb (8.6 kg) with optional transformers for either 100V or 230V AC operation
Fluid Interface
Two ¼ female NPT tapped threads provided at the Inlet and Outlet on the rear
panel.
Two ¼ male NPT connectors to ¼ swage tube fittings with two 3 (76mm) by ¼
OD copper tubes provided standard.
Two ¼ plastic Swagelock tube plugs are provided standard.
Options
Tube accessory kit with eight Swagelock quick disconnect connectors and 20 feet
(6 m) of polyurethane tubing ¼ OD by 3/16 ID.