Pioneer Q150 Respiratory Product User Manual


 
7
6. SPECIFICATIONS
Electrical Requirements
115VAC±10, 60Hz / 230VAC±10, 50Hz
Rated Current Input
3.5A (115VAC/60Hz); 2.0A (230VAC/50Hz)
Sound Level
45dB(A) Average
Operating Altitude
Do not use at altitude over 4000 meters.
Oxygen Concentration
90at 1~5LPM
* Stated performance maximum achieved after 60 min. of run time.
Maximum Oxygen
Delivery Pressure:
6±0.5psi. (40±4 kPa.)
Optional Oxygen Sensor
Oxygen Purity LED Indicator
Over 85(±3)
GREEN Indicator light
70(±3) to 85(±3)
YELLOW Indicator light
Below 70(±3)
RED Indicator light
Flow Range / Maximum
Recommended Flow
0-5 LPM / 5 LPM
Average Power
Consumption
320W
Pressure Relief Activation
250kPa±10
Filters:
Cabinet, Bacteria and Compressor Inlet
Alarms – (Visual/Audible)
High temperature w/ compressor – power shutdown
High / Low pressure alarm
Power loss alarm
Dimensions
68cm(26.8”) High × 45cm(17.7”) Wide × 33cm(13”) Deep
Circuit Breaker
& Fuse
(115VAC/60Hz)
Circuit Breaker: 8A
(230VAC/50Hz)
Circuit Breaker: 4A
Weight
28.2Kg (62lb)
Shipping Weight
31.8Kg (70lb)
Operating Ambient
Temperature
10 ~ 39