Omron Healthcare 510 Thermometer User Manual


 
13
Is the temperatures measured in the right ear different from that
measured in the left ear?
Among healthy people, there should be no significant difference in the
measurement results. Differences may be caused by the following reasons:
1) The infrared sensor is not inserted in the same way.
2) Measurement is not conducted by inserting the unit in a stable manner in
the same angle. Try to measure in either ear that can show the highest
measurement value constantly. In accordance with the “10 seconds mode,”
the unit detects the highest temperature. (Refer to paragraph 3.2).
The temperature shown is rather low.
1 The probe cover is dirty.
2 The infrared sensor is dirty.
3 You removed the unit from the ear before the measurement is finished.
4 The ear is cold. The temperature tends to indicate low when you use
an ice bag or an ice pack, or immediately after coming in from the
cold in winter.
5 The thermometer is not inserted deep enough in the ear.
6 The infrared sensor is not directed towards the eardrum.
Use the 10 seconds measurement mode (Refer to paragraph 3.2).
6 SPECIFICATIONS
Name: GentleTemp 510
®
Model: MC-510-E2
Voltage: 3V DC (lithium battery CR2032)
Power consumption: 0.05 W
Battery life: With a new battery approx. 5000 measurements
Sensor: Thermopile
Temperature indication: In 3 digits of +°C, every 0.1°F or °C
Precision: According to NEN-EN 12470- 5 Clinical
(Celsius) thermometers – Part 5 Performance of infra-red
ear thermometers (with maximum device)
34°C – 42.2°C: ±0.2°C
Based on the measurement of the standard black
body at the room temperature of 25°C (RH 50 %)
Precision: According to NEN-EN 12470- 5 Clinical
(Fahrenheit) thermometers – Part 5 Performance of infra-red
ear thermometers (with maximum device)
93.2°F – 108°F: ±0.3°C
Based on the measurement of the standard black
body at the room temperature of 77°F (RH 50 %)
Measurement range: 34°C - 42.2°C (93.2 - 108°F)
Weight: Approx. 50 g (1.8 oz) (including the battery)
External dimensions: Approx. 4.6 cm (B) X 9.3 cm (H) X 5.7 cm (D)
Operating environment: Ambient temperature; 10°C to 40°C (50°F to 104°F)
Relative humidity: 30 - 85%
Storage environment: Ambient temperature; -20°C-60°C (-4°F- 140°F)
Relative humidity: 10 - 95%
Accessories: Storage case, Instruction Manual, probe covers
(10 pieces)
A probe cover is already attached to the main unit.
Options: Probe cover exclusively for GentleTemp 510
®
series, Probe Cover MC-EP1
GentleTemp 510 English
13413 MC-510-E2_ver.6_MC-510-E2 10/04/2013 17:20 Page 13