Omron NE-U22V Respiratory Product User Manual


 
Troubleshooting
21
Troubleshooting Troubleshooting
The Power Indicator is illuminated, but the unit does not nebulize.
Where to Inspect How to correct
Is the Medication Bottle filled with medication?
Is the Mesh broken, stained,
or extremely soiled?
Has any medication or water collected on the
electrode of the Main Unit?
Fill with medication. (Refer to Page 12)
Clean the Mesh Cap by disinfecting.
(Refer to Page 18)
If the unit does not nebulize even after
carrying out the above-mentioned procedure,
or the Mesh is broken, replace the Mesh Cap
with a new one.
(Refer to Page 19)
Wipe off the stain. (Refer to Page 17)
Is the medication being supplied to the
Medication Bottle?
(Is the medication in contact with the Vibrator?)
Use the unit slightly slanting the Main Unit
(the side with the Button) towards yourself.
(Refer to Page 15)
Absorb the medication or water with
a soft cloth.
(Refer to Page 15)
Is the electrode of the Main Unit stained?