Omron NE-U22V Respiratory Product User Manual


 
How to Disinfect
18
How to Disinfect How to Disinfect the Unit
The Micro-Air
®
nebulizer can be cleaned and disinfected easily because the Medication
Bottle and the Mesh Cap can be removed from the Main Unit. The unit can
be disinfected with benzalkonium chloride.
1 Make 0.1% dilute solution using benzalkonium chloride, and soak the
Medication Bottle and Mesh Cap in the solution for 10 minutes.
2 Take the Medication Bottle and Mesh Cap out of solution
to rinse it with water.
3 Wipe off the remaining water for faster drying.
4 If using commercial cleaners, follow manufacturers instructions.
Caution
Do not use disinfection solution containing sodium hypochlorite.
The unit may not be able to be used after getting covered with rust.
5 The following parts can be disinfected:
Medication Bottle
Mesh Cap
Mask
Mouthpiece
Mouthpiece Adapter