Omron NE-U22V Respiratory Product User Manual


 
Troubleshooting
20
Troubleshooting Troubleshooting
If a problem occurs while you are using the unit, please check the following information before calling.
The nebulization rate is extremely low, or the unit does not nebulize at all.
Where to Inspect How to correct
Is the Battery Low Indicator turned on?
Is the Mesh dirty?
Replace the batteries with new ones.
(Refer to Page 10)
Where to Inspect
Are the batteries worn?
Is the AC Adapter securely plugged into
the electric outlet or the connection port
of Main Unit?
Is the capacity of the rechargeable batteries
(NiMH) low?
Is the Medication Bottle installed to the Main
Unit correctly?
Replace the batteries with new ones.
(Refer to Page 10)
Insert it correctly. (Refer to Page 11)
Recharge the batteries with the commercially
available recharger or replace them
with new ones.
Install it correctly. (Refer to Page 13)
The Power Indicator does not turn on and the unit does not nebulize.
Clean the Mesh Cap by disinfecting to remove
any stains. (Refer to Page 18)
If the trouble is not corrected by the above
mentioned procedure, replace the Mesh cap
with a new one.
(Refer to Page 19)
Are the polarities of batteries or rechargeable
batteries placed in the wrong direction?
Insert batteries or rechargeable batteries in the
correct alignment.
(Refer to Page 10)
How to correct