Omron HBP-1300 Blood Pressure Monitor User Manual


 
31
EN
Environmental Conditions
Non-Invasive Blood Pressure (NIBP)
* Comparison with auscultation method performed by a trained professional.
DIA determined by the auscultation method is “K5”.
Note: Specifications may be changed without prior notice.
This blood pressure monitor fulfils the requirements of the EC directive 93/42/EEC (Medical Device
Directive). It also conforms to the European standard EN 1060, Non-invasive Sphygmomanometers
Part 1: General Requirements and Part 3: Additional Requirements for Electromechanical Blood
Pressure Measuring Systems.
Operational temperature and
humidity
Temperature range: 10 to 40°C (50 to 104°F)
Humidity range:
30 to 85%RH (not condensed)
Atmospheric pressure: 700 to 1060hPa
Storage and transportation Temperature range: -20 to 60°C (-4 to 140°F)
Humidity range: 10 to 95%RH (not condensed)
Atmospheric pressure: 500 to 1060hPa
Measurement technology Oscillometric
Measurement method Dynamic Linear Deflation method
Pressure display range 0 to 300 mmHg
Pressure display accuracy Within ±3 mmHg
NIBP measurement range SYS 60 to 250 mmHg
DIA 40 to 200 mmHg
Pulse rate 40 to 200 /min
NIBP accuracy* Maximum mean error within ±5mmHg
Maximum standard deviation within 8mmHg
Pulse rate accuracy Within ±5 % of reading
Reference standard: EN1060-1:1995+A2:2009
EN1060-3:1997+A2:2009
ISO81060-1:2007