Omron HBP-1300 Blood Pressure Monitor User Manual


 
17
EN
Utility Mode
“Utility Mode” allows you to configure the “Auto Power Off” and “Pressure Accuracy Confirmation” setting.
1. Confirm that the device is switched off.
If the power is on, hold down the [START/STOP] button for at least 3 seconds to
turn off the power.
2. Hold down the [MODE] button until the “Auto Power Off”
setting screen appears.
The “Initial Inflation Pressure Value” setting screen appears, and changes to the
“Auto Power Off” setting screen appears.
3. Press the [ ] or [ ] button to change the auto power off
setting.
The “5 min.” or “10 min.” setting is entered.
When the unit is powered by the battery pack and is not used for the set time (“5 min.” or “10 min.”),
the power automatically turns off to save battery power.
If a mid-priority alarm other than a low battery error (E40 error) has occurred, the power does not
turn off automatically.
4.When you have completed the “Auto Power Off” setting, press the [MODE]
button.
The “Pressure Accuracy Confirmation” screen appears. “0” appears.
5. Check the Pressure Accuracy.
Add pressure externally as explained on page 24.
Compare with the displayed value and verify that there is no problem.
6. When you have completed “Pressure Accuracy Confirmation”, press the
[START/STOP] button.
The power turns off.
“5” or “10”
appears.
“0” appears.