Omron HBP-1300 Blood Pressure Monitor User Manual


 
28
List of Error Codes
The alarm lamp flashes when a medium-priority alarm occurs, and lights steadily when a low-priority
alarm occurs.
To clear an alarm, press any button.
Q If a low-priority alarm and a medium-priority alarm occur at the same time, the medium-priority alarm
is displayed.
If the alarms are the same priority level, the error code of the alarm that occurred first is displayed.
However, to prevent battery consumption when powered by battery, a low battery error (E40 error) is
always given priority.
Q Example: E2
SYSTEM
NIBP
Error
code
Priority Description Points to check
E9 Medium Internal hardware error Contact OMRON HEALTHCARE.
Error
code
Priority Description Points to check
E1 Low
The cuff tube is not connected Firmly connect the cuff tube.
Air is leaking from the cuff.
Replace with an OMRON GS CUFF that
does not leak.
E2 Low
Did not inflate properly because the arm
or body moved during measurement.
Have the patient not move the arm or body,
and repeat measurement.
Moved body or arm during
measurement, or talked.
Have the patient not talk or move, and
repeat measurement.
The cuff is not applied correctly. Correctly apply the cuff.
The sleeve is rolled up and is
compressing the arm.
Remove the garment and re-wrap the cuff.
Measurement time has exceeded
specified time.
Specified time: 165 seconds
The measurement time exceeds the
expected time, so the measurement was
ended in order to avoid patient discomfort.
There is a possibility that measurement is
being repeated over and over due to air
leaking from the cuff.