For digital probe matching within the instrument (version 2.1 only) additional
commands are used as follows:
CORRECTIONS A (turns OFF B corrections) Turns A on
B (turns OFF A corrections) Turns B on
AB (turns ON BOTH A & B corrections)
OFF (turns OFF BOTH A & B corrections)
Whatever is set by this command is retained during power-down.
LA Lists Ch A corrections
LB Lists Ch B corrections
CA Enter Ch A corrections
CB Enter Ch B corrections
Query Says “Not enabled”, “A”, “B”, “AB”
XX* Disables corrections
ENABLE* Enables corrections
IC Initialise corrections (i.e.Erase) – useful after installing software
*For these commands to be actioned the SPANss key must be pressed as the
command is typed.
Corrections entered as Follows
Type TM Sends
CA <CR> How Many Points
2 <CR> Entry=/0 Ohms
100,000 <CR> Deg C?
0.000 <CR> 0100.0000 Ohms 0000.0000 Deg C ok?
Y <CR> Entry=/1 Ohms
138.510 <GR> Deg C?
100.000 <CR> 0138.5100 Ohms 0100.0000 Deg C ok?
Y <CR> Done
This would repeat for more points.
Pressing N at the ok? Prompt will allow re-entry of the point.
Once entered, type LA OR LB TO SEE/CONFIRM CONTENTS.
Rules of entry.
1) Ohmic values must always increment from point to point
2) All values must be entered to THREE decimal places i.e:- .001,0.001,10.001 are all?
ok.