Omega Speaker Systems OS550 Series Thermometer User Manual


 
Appendix: How Infrared Thermometry Works
A
A-4
Optics Field of View
Accurate measurement of temperature via infrared means depends
strongly on the size of the object and the distance between the
thermometer and the object. All optical devices (e.g. cameras,
microscopes, infrared thermometers) have an angle of vision,
known as a field of view or FOV, within which they see all objects.
In particular, the thermometer will measure a fixed proportion
of
the energy radiated by all
objects within its FOV. The user must
guarantee that the distance between the thermometer and the object
is defined so that only that object fills the FOV of the instrument.
Referring to Figure A-3, Objects Xand Yare within the FOV of
the thermometer. The measured temperature would fall somewhere
between the actual temperatures of the two objects. In order to
measure the temperature of Object Xaccurately, Object Ywould
need to be removed. In order to measure the temperature of Object
Yaccurately, the user would need to move closer to Object Y
until it completely filled the FOV of the thermometer. Alternatively,
the user could measure the temperature of Object Ywith a
thermometer with a smaller FOV.
Figure A-3. Field of View of a Thermometer/Transmitter
The distance-to-spot size ratio (
D
S) defines the field of view (FOV).
Thus, a
D
S = 10 gives you approximately a 1' spot size at a distance of
10'. For accurate spot size values, refer to the Field of View diagrams
shown in Figures 3-3 through 3-8.