Omega Speaker Systems OS532E Thermometer User Manual


 
Using the Handheld Infrared Thermometer
2
2-28
2.3.15 Logging Temperature Data in Real Time
(OS523E, OS524E,OS534E)
The thermometer can log temperature data in real time. The
logged data is stored in the non-volatile memory, so
removing the batteries will not affect or erase the data. The
data is logged based on the data recording interval (PRN)
which can be set anywhere from 1 to 1999 seconds. The
thermometer can log up to 800 data points. Therefore, the
logging period can be anywhere from 13 minutes (1 second
recording interval) up to 18.5 days (1999 second recording
interval). To log temperature in real time:
1. Aim at the target and pull the trigger. Press the key
or pull the trigger twice to lock the trigger. The
icon will appear on the display.
2. Press the or keys to adjust the Emissivity
value for the target.
3. Press the key until the display mode
appears.
4. Set the data recording interval (seconds) by pressing the
or keys.
5. Press the key until the LOG display mode appears.
6. Press the key to start logging temperature data in
real time. The display will show LOG on, and the unit starts
logging data based on the recording interval set in the
display menu. Press the key again, and the
unit stops logging data. The display will show LOG off.
7. After logging data, you can turn off the thermometer by
double clicking the trigger button. The icon will
disappear and the unit turns off.
LOCK
LOCK
FUNC
F
C
FUNC
F
C
LOCK
°F
on
LCK