23
Function check
If you think that Innovo
®
is not working
properly, follow this procedure.
• Screw on a new NovoFine
®
needle as
described on page 9.
• Prime Innovo
®
as described on page 10-11.
• Put the outer needle cap over the needle.
• Dispense 20 units into the outer needle cap.
The insulin should fill the lower part of the
outer needle cap. If Innovo
®
has released
too much or too little insulin, repeat the test.
If it happens again, do not use your Innovo
®
.
Contact your supplier. Do not try to repair
a faulty Innovo
®
.