Mintek LV-T Pacemaker User Manual


 
Stratos LV/LV-T Technical Manual 153
Pulse Width
0.1; 0.2; 0.3; 0.4; 0.5…(0.25)…1.5 ms
(3 channels separately programmable)
Sensitivity
A 0.1... (0.1)...1.0… (0.1)…1.5… (0.5)...7.5 mV
RV 0.5... (0.5)...2.5... (0.5)...7.5 mV
First Chamber Paced
RV
VV Delay after Pace/Sense
5 ms
Ventricular Refractory Period
150… (25)…500 ms
Atrial Refractory Period
Auto; 225… (25)…775 ms
Atrial Far-field Protection
After Vp: 30… (10)…220 ms
After Vs: 30… (10)…200 ms
PMT Detection/Termination
Off; On
PMT VA Criterion
250… (10)…500
PMT Protection
Auto; 175… (25)…600 ms
PMT Protection after VES
400… (25)…600 ms
Mode Switching
Off; On