NOTE:
The attending physician must notify the BIOTRONIK Service
Center about which of these events he/she wishes to be
informed.
2.17.4.3 Patient Report
It is possible to trigger a transmission through magnet
application over the CRT-P. The attending physician must
inform the patient in detail about operating the device and about
the physical symptoms which would warrant a magnet
application by the patient.
CAUTION
Patient-Activated Report - The magnet effect must be
programmed “synchronous” if the [Patient Report] function is
activated.
2.17.5 Description of Transmitted Data
The following data are transmitted by the Home Monitoring
system, when activated. In addition to the medical data, the
serial number of the implant is also transmitted.
The Monitoring Interval
Type of Last Home Monitoring Message and Time of the
Transmission of the last Home Monitoring Message provide
information regarding the interval of Home Monitoring data.
Resynchronization Therapy
% Pacing in the Atrium / 24 hours [%]
Atrial intrinsic rhythm [%]
CRT - %ventricular pacing [%]
Ventricular paces (AV Delay expired) [%]
Ventricular paces (triggered by RV Sense) [%]