Mintek LV-T Pacemaker User Manual


 
130 Stratos LV/LV-T Technical Manual
6.2 Patient Data Memory
Individual patient data can be stored in the Stratos CRT-Ps.
For example, the following are stored:
Patient name
Patient index (how the patient name is coded)
Implantation date
Symptoms
Etiology
ECG indication
QRS width
LV ejection fraction
Lead polarity
6.3 Safe Program Settings
Activating the preset values for the Safe Program is a quick and
convenient way to provide VVI / SSI pacing at a high output
setting in urgent situations. Listed in Table 26
are the Safe
Program settings for Stratos CRT-Ps.
Table 26: Safe Program Settings
Parameter Safe Program Settings
Mode VVI
Pacing Rate 70 ppm
Amplitude 4.8 V (ventricle)
Pulse Width 1.0 ms
Sensitivity 2.5 mV
Ventricular Refractory Period 300 ms
Pacing Polarity Unipolar
Sensing Polarity Unipolar
Single Chamber Hysteresis OFF
Magnet Effect AUTO