Lifescan Blood Glucose Monitoring System Blood Glucose Meter User Manual


 
38
Understanding
error messages
Message What it means What to do
The meter has detected
that the temperature is
above the system
operating range. Do not
perform a test until the
meter and test strips
reach a temperature
within the operating
range of 43–111
°
F.
Repeat the test after the meter
and test strips have reached a
temperature within the
operating range.
The meter has detected
that the temperature is
below the system
operating range. Do not
perform a test until the
meter and test strips
reach a temperature
within the operating
range of 43–111
°
F.
Repeat the test after the meter
and test strips have reached a
temperature within the
operating range.
DCO_06397301A_OTUM_OB_EN.qxd 05/05/06 9:16 AM Page 38