Leica Biosystems Leica HER2 FISH System - 30 Test Instructions for Use TA9217 EN-CE-Rev_D 08/04/2013
English
Figure 1: Interpretation Guide
Leica HER2 FISH System - 30 Test Interpretation Guide
Count as 2 orange signals and 1 green signal
Do not count. Nuclei with no signals or with signals of only 1 color
should not be scored. Only score those nuclei with one or more
FISH signals of each color
Count as 1 orange signal and 2 green signals
Count as 3 orange signals and 2 green signals
Count as 5 orange signals and 4 green signals. 1 orange signal is
diuse and 1 green signal is diuse
Count as 4 orange signals and 2 green signals. 1 orange signal is
diuse. Count 2 signals that are the same size and separated by
a distance equal or less than the diameter of the signal, as one
signal
Count as 5 orange signals and 3 green signals. 1 orange signal is
diuse
Count as 2 orange signals and 2 green signals. 1 orange signal and
1 green signal are overlapping
Do not count. The nuclei are overlapping. It is too dicult to tell
which nuclei the signals are located in
Count as approximately 16 orange signals and 2 green signals