Leica RM2125 Microscope & Magnifier User Manual


 
Leica RM2125 Troubleshooting Guide
Problem
1. Thick and thin sections
(continued)
a) Partial sections
b) Shorter sections in ribbon
c) Section thickness varies from one section to another
d) Skipping sections
e) Varied thickness in one individual section
Cause(s) / Corrective Action
VI) Backlash in specimen advance system.
l) Test for backlash by trying to rotate the object head clockwise and counter clockwise. If there is play in the
system, schedule a service call for the internal adjustments of the specimen advance system.
2
back to Table of Contents