Leica RM2125 Microscope & Magnifier User Manual


 
Leica RM2125 Troubleshooting Guide
Problem
10. Microtome makes "clunking noise" when sectioning
Cause(s) / Corrective Action
I) Linear bearings in cage are at their lower limits.
a) Linear bearings need to be cleaned and possibly re-greased.
II) Alignment of bearing cage is incorrect.
b)
Schedule a service call.