Leica RM2125 Microscope & Magnifier User Manual


 
Leica RM2125 Troubleshooting Guide
Problem
5. No specimen advance (no section produced)
Cause(s) / Corrective Action
I) Specimen reached the front feed limit.
a) Turn the coarse feed wheel in the appropriate direction to move the specimen towards the rear limit.
II) User may accidently be holding the coarse feed wheel while sectioning.
III) The coarse feed wheel cannot rotate freely.
b) Remove any obstruction.
IV) Internal mechanism (gear) may be faulty.
c) Schedule a service call for replacement of the coarse advance mechanism.
8