Leica MZ16 F Microscope & Magnifier User Manual


 
Leica MZ16 F – Safety concept 9
1.3 Light source: safety instructions
Precautions by the manufacturer
UV protection screen in front of the specimen
plane prevents the user from looking directly
into the UV rays.
Simple filter holders (that have only two open-
ings for the observation beam paths) in the avail-
able positions of the rapid filter changer prevent
direct UV radiation from reaching the eyes.
UV shutter (which can be activated/deacti-
vated) prevents direct UV radiation from reach-
ing the eyes.
UV filters in the observation beam path protect
the eyes.
The stray-light protection on the lamp housing
prevents irradiation of the hands.
Warning
UV radiation can harm the eyes.
Therefore:
Never look into the light spot on the specimen
plane without a UV protection screen.
Always fill all of the filter positions with filter
holders.
Fill empty filter positions with the simple filter
holders (which have only two openings for the
observation beam paths).
Do not select a white, strongly reflective
background for the stage.
Never open the lamp housing while the lamp is
switched on.
Risk of explosion, UV exposure, blinding!
Supply unit
Always unplug the supply unit from the power
supply:
when installing and disassembling the lamp
housing
before opening the lamp housing
when replacing the high-pressure mercury
burner and other parts, such as the heat-
absorbing filter or the collector
during maintenance work on the supply unit.
Lamp housing
The Leica MZ16 F may be used only together with
the Leica 106 z lamp housing for the 100W or 50W
mercury burner, supply unit and stray-light protec-
tion.
Leica Microsystems accepts no liability for
damages resulting from the use of lamp housings
from other manufacturers.
Never open the lamp housing while the lamp is
switched on. Unplug the supply unit from the
power supply:
Risk of explosion, UV exposure, blinding!
Before opening the lamp housing, allow it to
cool off for at least 15 minutes.
Danger of explosion, danger of burns.
Never cover the air duct on the lamp housing.
Danger of fire!
Use the stray-light protection.
Danger to eyes from UV radiation.