5.3 Rapid filter changer
The rapid filter changer must be equipped with
four filter holders at all times.
If you are using fewer than four filter hold-
ers with fluorescence filters, always insert the
included simple filter holders into the available
positions. Otherwise there is a risk of danger to
the eyes from direct UV radiation (see page 16).
If at all possible, avoid touching the filters
in the filter sets in order to avoid fingerprints.
If you touch them, clean the filters using a soft,
lint-free cloth and pure alcohol.
̈ Insert a filter set into the rapid filter changer
such that the label (e.g. GFP) is upright and leg-
ible on the right side.
Ensure that the contour of the filter set matches
the contour of the microscope.
̈ Rotate the filter set by hand until it latches into
place and insert the next filter set.
Insert a total of four filter sets or the included sim-
ple filter holders into the empty positions.
5.2 Commissioning the fluorescence
system
After switching on the high-pressure mer-
cury lamp, you must adjust the burner immedi-
ately (see User Manual M2-216-1). Do not switch
on the supply unit until you can safely operate the
Leica MZ16 F.
Follow all safety instructions when
handling mercury lamps and the lamp housing!
̈ Switch on the supply unit (see User Manual
M2-216-1).
Wait for 2 to 3 minutes.
̈ Adjust the burner in accordance with User
Manual M2-216-1.
For observation with fluorescence light,
always position the aperture protection correctly
to protect the observer (page 16).
1
2
Fig. 9 Filter changer, equipped with filter holders
1 Filter sets
2 Filter slide for filter holder with individual filters