Konica Minolta PULSOX-300i Oxygen Equipment User Manual


 
9
Foreword
Differencefromthe SaO
2
value to displayed value
<Table 1> <Table 2>
SaO
2
HbCO
SaO
2
Methemoglobin
1% 5% 10% 1% 5% 10%
50 50
70 70
90 90
100 100
Measured value is always
lower than theactual value.
Displayed value is sometimes
higher than theactual value if
SaO
2
is around 50%.
-0.1 -0.7 -1.5 +0.2 +1.3 +3.2
-0.2 -0.8 -1.6 -1.5 -6.0 -9.6
-0.1 -0.7 -1.5 -0.6 -2.3 -3.2
-0.2 -0.8 -1.7 -1.8 -7.5 -12.2
When inuenced electromagnetically by other electronic de-
vices (e.g. near electrical appliances like TV, medial device)
When a mobile telephone is used during measurement
When influenced by abnormal hemoglobin like carbon
monoxide hemoglobin (HbCO) and methemoglobin
When pigments such as cardiogreen, intravascular dyes
and indocyanine greenare present in the blood
When nger nails are polished
Tables 1 and 2 below show the errors which may occur due to
these hemoglobins.
Do not immerse the probe, the extension cable or the con-
necter of PULSOX-300i in disinfectant. Doing so may cause
a malfunction.
When using the Extension Cable EC-300, t and secure ca-
bles so they are not too loose or too tight.
Before connecting to a personal computer, remove the Pulsox
from the patient.