SECTION 7 —TROUBLESHOOTING GUIDE
Platinum™Series 32 Part No 1145728
Alarm:
RAPID
Beep....Beep...
Beep...Beep
LOW FLOW ALARM
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at 0.5
L/min. for Platinum 5 and
1 L/min. for Platinum 10
or less.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back On.
1b. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is below
0.5 L/min., use the Pediatric
Flowmeter IRCPF16 (Only
with Platinum 5 and XL).
SYMPTOM: PROBABLE CAUSE: SOLUTION: