Invacare XL Oxygen Equipment User Manual


 
SECTION 7 —TROUBLESHOOTING GUIDE
Platinum™Series 30 Part No 1145728
SECTION 7 — TROUBLESHOOTING
GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
Alarm: Main Power Loss: 1. Insert plug into outlet.
Short beeps, long
pause
1. Power cord NOT
plugged in.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
Concentrator not
operating,
power switch On.
2. No power at outlet. 3. Call Service Provider.
Beep......Beep...... 3. Tripped circuit
breaker.
Alarm:
Continuous
Concentrator not
operating, Power
Switch On.
Beeeeeeeep....
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required
1a. Remove and clean cabinet
filters.
1b. Move oxygen
concentrator at least three
inches away from walls,
draperies or furniture.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.