Invacare XL Oxygen Equipment User Manual


 
SECTION 5 —OPERATING INSTRUCTIONS
Part No 1145728 23 Platinum™Series
FIGURE 5.4 Attaching the Oxygen Tubing
ƽ WARNING
For optimum performance, Invacare recommends that
each concentrator be on and running for a minimum of 30
minutes at a time. Shorter periods of operation may
reduce maximum product life.
Power Switch
NOTE: For this procedure, refer to FIGURE 5.5 on page 23.
1. Press power switch to on position. All the panel lights and
the audible alarm will come on for one second, indicating
that the unit is functioning properly. After one second,
only the GREEN system ok/power light will stay on.
FIGURE 5.5 Power Switch
Oxygen
Tubing
Humidifier Outlet
Oxygen Outlet
Connector
On/Off (I/O)
Power Switch