Invacare XL Oxygen Equipment User Manual


 
SECTION 4 —SPECIFICATIONS
Platinum™Series 14 Part No 1145728
SECTION 4 — SPECIFICATIONS
*NOTE: Stated concentration levels achieved after initial warm-up
period (approximately 30 minutes).
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Rated Current Input: Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Sound Level: Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Altitude: Platinum 5 and XL - Up to 6,000 ft. (1828
meters) above sea level without degradation
of concentration levels. From 6,000 ft. (1828
meters) to 13,129 ft. (4000 meters) below
90% efficiency
Platinum 10 - Up to 4,000 ft. (1230 meters)
above sea level without degradation of
concentration levels. From 4,000 ft. (1230
meters) to 13,129 ft. (4000 meters) below
90% efficiency
*Oxygen Output
Concentration Levels:
All 5LXO2/5LX models 95.6% to 87% at 0.5
to 5 L/min. and all 10LXO2/10LX models
95.6% to 87% at 2 L to 10 L/min.
Maximum Outlet Pressure: Platinum 5 and XL: 5 psi ± 0.5 psi. (34.5 kPa
± 3.45 kPa.)
Platinum 10: 9 psi ± 0.5 psi. (62.0 kPa ± 3.45
kPa.)