Invacare IRC5P Oxygen Equipment User Manual


 
SECTION 23—LEAK TEST
Perfecto
2
76 Part No 1148070
SECTION 23—LEAK TEST
ƽ DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
Leak Test
NOTE:Forthisprocedure,refertoFIGURE 23.1onpage 77.
1. Unplugunit.
2. Removecabinet.RefertoRemovingCabinet
onpage 23.
3. Pluginunit.
4. Turntheconcentratoron().
5. Letconcentratorrunfor20minutes.
ƽ WARNING
Applysmallamountofleaktestsolutiontofittingsonly.Avoidallothercomponents.DO
NOTleaktest4wayvalvebody.Carefullyleaktestforpressureleaksaroundsievebed
hosesat4wayvalve.DONOTallowleaktestsolutiontoenterinto4wayvalveand
system.
6. Withʺleaktest”solutionorequivalentsoapywater,leaktestthefollowing:
A. Thetwosievebedhoseconnectionatthe4wayvalve/manifoldassembly
(Detail”A”).
B. Leaktestthehosefittingoncenterportof4wayvalve/manifoldassembly.
C. Hoseconnectionstotopandbottomfittingsofsievebeds.
D. P.E.valvehoseconnections.
E. Hose
fittingsatproducttankcapandcheckvalves(Detail“B”).
F. Bothfittingsonheatexchangerassembly(Detail“C”).
7. Replaceanytubingthatappearscracked,worn,etc.
8. Reinstallcabinet.RefertoRemovingCabinet
onpage 23.
9. RefertoTypicalProductParametersonpage 8forconcentrationlevels.Iflow
concentrationsareobserved,refertoTroubleshootingonpage 16.
I/