Invacare IRC5P Oxygen Equipment User Manual


 
SECTION 16—TRANSFORMER
Part No 1148070 61 Perfecto
2
SECTION 16—TRANSFORMER
Replacing the Transformer
ƽ DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:Forthisprocedure,refertoFIGURE 16.1onpage 61.
1. Unplugunit.
2. Removecabinet.RefertoRemovingCabinet
onpage 23.
3. Locatethetransformerlocatedonthesoundbox.
4. Codeallwiresbeforedisconnectingforproperreconnections.
5. Removethetwomountingscrewsthatsecurethetransformertothesoundbox.
6. Noteorientationoftransformeronsoundboxbeforeremovingforproper
reinstallation
7. Removetheexistingtransformeranddiscard.
8. Installnewtransformerinorientation
notedinSTEP7.
9. ConnectallwiresasnotedfromSTEP4.
10. Reinstallcabinet.RefertoCabinetonpage 23.
FIGURE 16.1 Replacing the Transformer
Transformer
Assembly
Location of
Transformer
DETAIL “A”
Mounting
Screws