Invacare IRC5P Oxygen Equipment User Manual


 
SECTION 7—COMPRESSOR
Part No 1148070 35 Perfecto
2
CAUTION
DONOTlubricateoruseoilonanymovingparts.Thecompressoreccentricusesa
precisionsealedbearing.Therefore,additionallubricationisnotnecessary.
Removing Compressor and Compressor Heads
Compressor
1. Unplugconcentratorandcarefullyremovecompressorfromconcentrator.Referto
ReplacingCompressorAssemblyonpage 31.
Compressor Head Plate
NOTE:Forthisprocedure,refertoFIGURE 7.3.
1. Cleanloosedirtfromtheoutsideofthe
compressor.
2. UsetheTorxT25torquewrenchand
loosentheeightheadscrews.
3. Noteheadplacementbeforeremoving
compressorheadplate.
4. Removethecompressorheadplate.
Valve Plates
NOTE:Forthisprocedure,referto
FIGURE 7.3.
1. Notethepositionoftheplatesbefore
removing.
2. Liftoffthevalveplateononeside.
3. Repeatfortheotherside.
NOTE:Notethepositionofthetabonvalve
plateforreassembly.
NOTE:Toavoidconfusion,ONLYserviceone
endofthecompressoratatime.
FIGURE 7.3 Removing Compressor and
Compressor Heads - Rebuilding the
Compressor - Compressor - Compressor
Head Plate - Valve Plates
Head
Screws
Head Screws
Compressor
Head
Plate
Valve
Plate
Valve
Plate
Compressor